Determination of Size and Concentration of Particles in Oils
The concentration and size distribution of particles in an oil sample may be measured using the Coulter Principle also known as the Electrical Sensing Zone (ESZ) method. A suitable organic electrolyte solution is chosen to perform the analysis.The sample is prepared by dissolving a certain volume of oil in the electrolyte and analyzing it using a Beckman Coulter Multisizer 3 to determine the size distribution and concentration for the particles present in the oil. If the oil being tested is not soluble in the electrolyte,it is necessary an intermediate step involving the use of a solvent capable of dissolving the oil and then from this solution a sample is taken and dispersed into the electrolyte. A blank run is subtracted from the sample. The results are then reported as number of particles per milliliter for the desired size range.
SIGNIFICANCE
The International Standard ISO 4406 defining the Solid Contaminant Code for Hydraulic Fluids and Lubricants classifies the oils according to their particle content in the following ranges:
Particles /ml
> 4 µm
>6 µm
> 14 µm
This is the most widely used range to analyze the particle content in oils. A 100 µm aperture tube (2µm to 60 µm range) will be suitable. For a different analysis range other aperture tubes may be used.
3. Instrument Set up and Calibration
Select the size of the aperture tube appropriate for the analysis size range. The linear dynamic range for any aperture is 2% to 60% of its size, i.e. a 100 µm aperture tube will be capable of analyze the particle concentration and size distribution from 2 µm to 60 µm. Set up and calibrate the instrument according to the Multisizer 3 Operator’s Manual. For determining particle concentration the control mode for the instrument must be Volumetric Mode. Select any volume up to 2000 µL; the volume required for the analysis has to be determined experimentally depending upon the state of the oil and the length of the analysis.
PROCEDURE
4.1 Preparation of electrolyte solution
A 2% solution of NH4SCN (Ammonium Thiocyanate) in Isopropyl alcohol (IPA) is suitable for the majority of oils. Prepare the electrolyte by dissolving 20 g NH4SCN in 1.0 L of Isopropyl alcohol. Filter the solution using a 0.45 µm alcohol-compatible membrane filter.
4.2 Sample preparation
If the oil being tested is not soluble in the electrolyte, it is necessary an intermediate step involving the use of a solvent capable of dissolving the oil and then from this solution a sample is taken and dispersed into the electrolyte.
4.2.1 Oils soluble in Isopropyl alcohol
Measure exactly 20 mL of electrolyte solution into a 20 mL Accuvette® II. Pipette 2.0 mL of oil into the electrolyte, this quantity may be different according to the state of the oil. Stir gently to dissolve thoroughly without creating bubbles.
4.2.2 Oils not soluble in Isopropyl alcohol
Measure exactly 10 mL of Methyl Isobutyl Ketone (MIK) into an appropriate size glass flask. Pipette 2.0 mL of oil into the flask containing the MIK, this quantity may be different according to the state of the oil. Stir gently to dissolve thoroughly without creating bubbles. Measure exactly 20 mL of electrolyte solution into a 20mL Accuvette® II. Pipette 2.0 mL of the MIK-oil mixture into the electrolyte. Stir gently to dissolve thoroughly without creating bubbles.
4.3 Entering the sample information in the Multisizer 3 Software
Enter the required sample information in the software, i.e. dilution factor, analytical volume and volume of oil used to prepare the sample. By entering the sample information, the software calculates the concentration of particles in the original oil.

4.4 Sample Analysis
Prepare a blank by following the same procedure used according to the kind of oil to be analyzed but without adding the oil. Run the blank and set the software to automatically subtract it from all the subsequent runs. Run the oil samples.
REPORTING THE RESULTS
Results can be represented as the entire size distribution, total number of particles or number of particles above, below or within selected size categories in the range of the size distribution by using the Interpolation feature in the Multisizer 3 Software.
Helpful Links
-
Material de lectura
-
Notas de aplicación
- 17-Marker, 18-Color Human Blood Phenotyping Made Easy with Flow Cytometry
- 21 CFR Part 11 Data Integrity for On-line WFI Instruments
- 8011+ Reporting Standards Feature and Synopsis
- Achieving Compliant Batch Release – Sterile Parenteral Quality Control
- Air Particle Monitoring ISO 21501-4 Impact
- An Analytical Revolution: Introducing the Next Generation Optima AUC
- Analyzing Mussel/Mollusk Propagation using the Multisizer 4e Coulter Counter
- Automated 3D Cell Culture and Screening by Imaging and Flow Cytometry
- Automated Cell Plating and Growth Assays
- Automated Cell Transfection and Reporter Gene Assay
- Mediciones automatizadas de concentración y viabilidad de las células sanguíneas del cordón umbilical utilizando el VI-CELL XR
- Automated Genomic Sample Prep-Ampure XP PCR
- Automated Genomic Sample Prep-RNAdvance
- Automated Genomic Sample Prep-Sample Quality Control
- Automated salt-assisted liquid-liquid extraction
- Automated Sample Preparation for the Monitoring of Pharmaceutical and Illicit Drugs by LC-MS/MS
- Métodos de transfección automatizados
- Automated XTT Assay for Cell Viability Analysis
- Automatización de un gradiente de densidad lineal para purificación de un complejo proteína-ligandos
- Automatización del control de calidad biofarmacéutica para reducir costes y mejorar la integridad de los datos
- Automatización de los ensayos Bradford
- Automatización del desarrollo de líneas celulares
- Automating Cell Line Development
- Automatización de procesos basados en células
- Automatización del flujo de trabajo de desarrollo de líneas celulares
- La nueva centrífuga Avanti J-15 mejora la protección de las muestras y maximiza la recuperación de muestras
- El nuevo perfil de desaceleración de ahorro de tiempo de la centrífuga Avanti J-15 mejora la eficiencia del flujo de trabajo
- Avanti JXN Protein Purification Workflow
- Evite los errores al automatizar el recuento de viabilidad celular para el control de calidad biofarmacéutico
- Beer, Evaluation of Final Product and Filtration Efficiency
- La preparación automatizada de muestras genómicas de Biomek acelera la investigación
- Biomek Automated NGS Solutions acelera la investigación genómica
- Preparation and purification of carbon nanotubes using an ultracentrifuge and automatic dispensing apparatus, and analysis using an analytical centrifuge system
- Recuento de células del analizador de viabilidad celular Vi–Cell BLU
- Cell Line Development – Data Handling
- Cell Line Development – Hit Picking
- Cell Line Development – Limiting Dilution
- Cell Line Development – Selection and Enrichment
- Cambios en las reclasificaciones de BPF de sala limpia en vigor
- Caracterizando la insulina como producto biofarmacéutico usando el ultracentrifugado analítico
- Clean Cabinet Air Particle Evaluation
- Monitorización ambiental rutinaria de sala limpia – Guía de la FDA sobre 21 CFR Parte 11 Integridad de los datos
- Comparación de la calidad de los datos y la resolución óptica del Optima AUC de nueva generación con el ProteomeLab de eficacia demostrada en un sistema de proteínas modelo
- Realización de la prueba de recuperación de sala limpia ISO 14644-3 con el MET ONE 3400
- Considerations of Cell Counting Analysis when using Different Types of Cells
- Consistent Cell Maintenance and Plating through Automation
- Normas de control y recomendaciones de métodos para el LS 13 320 XR
- Eficacia del recuento: Contadores de partículas de aire MET ONE y cumplimiento de ISO-21501
- Critical Particle Size Distribution for Cement using Laser Diffraction
- CytoFLEX
- Detección de humedad en fluidos hidráulicos, aceite y combustibles
- Determination of Size and Concentration of Particles in Oils
- El aislamiento eficiente de ácidos nucleicos sin kit utiliza una combinación de métodos de separación mediante precipitación y centrifugación
- dsDNA Quantification with the Echo 525 Liquid Handler for Miniaturized Reaction Volumes, Reduced Sample Input, and Cost Savings
- Configuración de compensación para reactivos DURAClone de alto contenido
- Echo System-Enhanced SMART-Seq v2 for RNA Sequencing
- Efficient Factorial Optimization of Transfection Conditions
- Enhancing Vaccine Development and Production
- Enumeration And Size Distribution Of Yeast Cells In The Brewing Industry
- Farmacopea europea EP 2.2.44 y carbono orgánico total
- Evaluación del rendimiento de instrumento a instrumento del analizador de viabilidad celular Vi-CELL BLU
- FBS mermado de exosomas usando la centrifugación Beckman Coulter: La opción rentable y uniforme
- Flexible ELISA automation with the Biomek i5 Workstation
- Automatización completa del flujo de trabajo SISCAPA® utilizando una estación de trabajo de automatización de laboratorio Biomek NXP
- Análisis celular totalmente automatizado mediante citometría de flujo
- Get Control in GMP Environments
- g-Max: Added Capabilities to Beckman Coulter's versatile Ultracentrifuge Line
- Grading of nanocellulose using a centrifuge
- A method of grading nanoparticles using ultracentrifugation in order to determine the accurate particle diameter
- Grading of pigment ink and measurement of particle diameter using ultracentrifugation / dynamic light scattering
- HIAC Industrial: nuestra solución general para la prueba de fluidos de potencia para todas las aplicaciones
- HIAC PODS+ Online Mode & Filter Cart Mode
- HIAC PODS+ versus Parker ACM-20 Performance comparison
- A complete workflow for high-throughput isolation of DNA and RNA from FFPE samples using Formapure XL Total on the KingFisher™ Sample Purification System: an application for robust and scalable cancer research and biomarker discovery
- A Highly Consistent BCA Assay on Biomek i-Series
- A Highly Consistent Bradford Assay on Biomek i-Series
- A Highly Consistent Lowry Method on Biomek i-Series
- Highly Reproducible Automated Proteomics Sample Preparation on Biomek i-Series
- Desarrollo de líneas celulares: selección de coincidencias
- Cómo utilizar la dispersión lateral violeta (V-SSC) para detectar nanopartículas en el citómetro de flujo CytoFLEX
- Cómo la dispersión lateral violeta permite la detección de nanopartículas
- HRLD Recommended Volume Setting
- Hydraulic Oil Measurement
- ICH Q2: el desafío de medir el carbono orgánico total en sistemas de agua farmacéutica modernos
- ICH Q2 – The Challenge of Measuring Total Organic Carbon in Modern Pharmaceutical Water Systems
- Importancia de la medición del COT en el API (agua para inyección) a la luz del cambio de la Farmacopea Europea
- Dependencia con la temperatura del radio hidrodinámico de una proteína intrínsecamente desordenada medida en la ultracentrífuga analítica Optima AUC.
- Isolation of cell-free DNA (cfDNA) from plasma using Apostle MiniMax™ High Efficiency cfDNA Isolation kit— comparison of fully automated, semi-automated and manual workflow processing
- Problemas con las pruebas de contaminación de los combustibles para aviones
- Aprovechamiento del Vi-CELL MetAfLEX para monitorizar la actividad metabólica celular
- Linealidad de BSA utilizando ópticas de absorbancia y interferencia
- Long Life Lasers
- LS 13 320 XR: Sample Preparation - How to measure success
- Particle Size Analysis Simple, Effective and Precise
- Matching Cell Counts between Vi–CELL XR and Vi–CELL BLU
- MET ONE Sensor Verification
- Metal colloid purification and concentration using ultracentrifugation
- Separation and purification of metal nanorods using density gradient centrifugation
- Método para determinar el ajuste del parámetro de tipo de célula para que concuerde con el Vi CELL XR heredado
- Minimal Sample to Sample Carry Over with the HIAC 8011+
- Minimizar la variabilidad del proceso en la producción de agua potable embotellada
- Tendencias modernas en la monitorización de partículas no viables durante el procesamiento aséptico
- Ultracentrifugado analítico de múltiples longitudes de onda de albúmina de suero humano complejada con porfirina
- Particle diameter measurement of a nanoparticle composite - Using density gradient ultracentrifugation and dynamic light scattering
- Qué se hace ahora con el ACFTD discontinuado
- Optimizing the HIAC 8011+ Particle Counter for Analyzing Viscous Fluids
- Optimizing the Multisizer 4e Coutler Counter for use with Small Apertures
- Optimización del flujo de trabajo Eficiencia de la monitorización ambiental rutinaria de sala limpia
- Recuento de partículas en aplicaciones mineras
- Pruebas de partículas en líneas de gas a alta presión de sala limpia en conformidad con ISO 14644 fáciles de realizar con las calibraciones de gas del MET ONE 3400
- Pharma Manufacturing Environmental Monitoring
- Pharma Manufacturing Paperless Monitoring
- Principles of Continuous Flow Centrifugation
- Protein purification workflow
- Background Subtraction
- Calibrating the QbD1200 TOC Analyzer
- Detection Limit
- Inorganic Carbon Removal
- JP SDBS Validation
- Method Overview
- Overload Recovery
- QbD1200 Preparing Reagent Solution
- USP System Suitability
- Registros electrónicos de control de calidad para el cumplimiento de 21 CFR parte 11
- Using the Coulter Principle to Quantify Particles in an Electrolytic Solution for Copper Acid Plating
- RCC A standardized automated approach for exosome isolation and characterization
- RCC Avanti J 15 Series Maximizing Sample Protection and Sample Recovery
- RCC Avanti J 15 Series Time Savings and Workflow Efficiency
- Investigaciones de causa raíz para sistemas de agua farmacéutica
- Purificación de plásmidos escalable con CosMCPrep
- Comparación de especificaciones del Vi–CELL XR y el Vi–CELL BLU
- Especificación de la monitorización de partículas no viables para procesamiento aséptico
- Un enfoque estandarizado y automatizado para el aislamiento y la caracterización de exosomas utilizando instrumentos Beckman Coulter
- Cambio de prueba de aceite a agua y viceversa utilizando el HIAC 8011+ y el HIAC PODS+
- Análisis avanzado de subconjuntos de linfocitos T humanos en el citómetro de flujo CytoFLEX utilizando un reactivo seco DURAClone de 13 colores en tubo
- Using k-Factor to Compare Rotor Efficiency
- USP 787 Small Volume Testing
- Validación de los analizadores de carbono orgánico total en línea para pruebas de liberación utilizando ICH Q2
- Descontaminación con peróxido de hidrógeno vaporizado del instrumento Vi–CELL BLU
- Vi CELL BLU FAST Mode Option
- Cumplimiento reglamentario del Vi-CELL BLUE: 21 CFR parte 11
- A workflow for medium-throughput isolation of cfDNA from plasma samples using Apostle MiniMax™ on the KingFisher™ Technology
- Folletos, volantes y fichas técnicas
- Catálogos
- Folletos
-
Pósteres
- Aplicaciones de ultracentrifugación en la purificación y caracterización de biomoléculas
- Automatización de la extracción de ADN genómico de sangre completa y suero con GenFind V3 en la estación de trabajo genómica híbrida Biomek i7
- Extracción automatizada de ADN genómico de grandes volúmenes de sangre completa
- Aislamiento y separación de ADN y ARN de una única muestra de cultivo celular o tejido
- Correlación entre mutaciones encontradas en tejido tumoral FFPE y muestras de ADNlc emparejadas
- Un método escalable y automatizable para la extracción de ADNlc
- Extracción simultánea de ADN y ARN de tejido fijado en formol e incrustado en parafina (FFPE)
- AMP 2019: Correlation Between Mutations Found in FFPE Tumor Tissue and Paired cfDNA Samples
- Automated library preparation for the MCI Advantage Cancer Panel at Miami Cancer Institute utilizing the Beckman Coulter Biomek i5 Span-8 NGS Workstation
- Automating Cell Line Development for Biologics
- Cellular Challenges: Taking an Aim at Cancer
- Cell-Line Engineering
- Characterizing the Light-Scatter Sensitivity of the CytoFLEX Flow Cytometer
- ASHG 2019: Comparison between Mutation Profiles of Paired Whole Blood and cfDNA Samples
- ASHG 2019: Correlation Between Mutations Found in FFPE Tumor Tissue and Paired cfDNA Samples
- Mastering Cell Counting
- Preparing a CytoFLEX for Nanoscale Flow Cytometry
- A Prototype CytoFLEX for High-Sensitivity, Multiparametric Nanoparticle Analysis
- Instrucciones del producto
- Protocolos
-
Estudios de casos
- Adenoviral Vectors Preparation
- Algae Biofuel Production
- Autophagy
- B Cell Research
- Cardiovascular Disease Research
- Cell Marker Analysis
- Choosing a Tabletop Centrifuge
- Collagen Disease Treatment
- Controlling Immune Response
- Creating Therapeutic Agents
- DNA Extraction from FFPE Tissue
- English Safety Seminar
- Equipment Management
- Exosome Purification Separation
- Future of Fishing Immune Research
- Hematopoietic Tumor Cells
- iPS Cell Research
- Fundamentos de la purificación de proteínas de membrana para cristalografía de rayos X
- Organelles Simple Fractionation
- Particle Interaction
- Retinal Cell Regeneration
- Sedimentary Geology
- Severe Liver Disease Treatment
- Treating Cirrhosis
- University Equipment Management
- Virus Purification Fundamentals
-
Informes técnicos
- La centrifugación es una solución de flujo de trabajo completa para la purificación de proteínas y la cuantificación de la agregación proteica
- AUC Insights - Analysis of Protein-Protein-Interactions by Analytical Ultracentrifugation
- AUC Insights - Assessing the quality of adeno-associated virus gene therapy vectors by sedimentation velocity analysis
- AUC Insights - Sample concentration in the Analytical Ultracentrifuge AUC and the relevance of AUC data for the mass of complexes, aggregation content and association constants
- Hydraulic Particle Counter Sample Preparation
- Libros electrónicos
- Entrevistas
-
Notas de aplicación